Zeroing HLS circuit

The HLS circuit pressures are only zeroed at the time of circuit priming before the tubing is filled with fluid. There is no need to routinely re-zero at the time of use or thereafter.

Circumstances where a primed HLS may need to be re-zeroed

  • The zeroing step was missed during the priming procedure
  • The primed circuit has been placed on a different console and implausible pressure readings are obtained (If the readings are no different on a new console re-zeroing is NOT required)
  • There is concern that the pressure readings are erroneous on a current patient

Note: when the circuit is prepared for use the pressure readings reflect the hydrostatic pressures from the priming bag and should NOT trigger re-zeroing (e.g. with a typical elevation of the priming bag to about 80cm above the sensor the pressure reading will be around 60mmHg).

Zeroing pressures on a primed HLS circuit

There are two scenarios

The circuit is not connected to a patient

The priming bag has to be lowered to the level of the console to eliminate any hydrostatic pressure. Set RPM to 0 to ensure no flow in the circuit then zero the pressures.

The circuit is running on a patient

Discuss with the clinician in charge regarding the need to zero.

  • The patient needs to be at the same level as the console pressure transducers
  • Apply a clamp on the return limp first (to eliminate the systemic blood pressure)
  • Turn RPM to zero
  • Clamp the access limb, then proceed to zero the pressure

The clamp on the access side is somewhat optional since you will still zero to the patient’s venous pressure.

Note: Re-zero refers to all 3 pressures (pre/ inter/ post). In the process of zeroing there is a checkbox to confirm that the tubing is not filled with fluid. This can be confirmed, the difference between fluid-filled is small (the introduced error is about 10mmHg or more precisely all pressures will underread by the central venous pressure) as long as the patient is supine level with the transducers and no additional hydrostatic pressure applied to the sensor.

Flow Signal Zeroing on the Cardiohelp or Rotaflow II platform may be performed if required

  • Clamp circuit either side of the flow sensor (ensure the direction of flow sensor arrow is correct)
  • Turn RPM to zero
  • Select the flow monitoring screen on the touch screen panel and then press the zero indicator button once until zero flow is confirmed on the flow panel
  • Pump speed is set to approximately 1500 RPM and the clamps are released before gradually increasing RPM until the desired flow is reached