These sections will demonstrate the start-up of the CardioHelp and Rotaflow II consoles with a pre-primed circuit. Circuit priming guides can be found here (HLS/CardioHelp) (PLS/Roatflow II)
Cardiohelp start-up
Youtube video blocked
This YouTube video has been blocked from loading as you have not given consent to the associated cookies.
Priming HLS circuit can be found here.
Emergency drive unit
An emergency drive unit (“hand crank”) must be positioned on the trolley such that the centrifugal pump can be transferred immediately in the event of either console or pump head failure.

The emergency hand crank is freely accessible in proximity to the console, the heater connection may need to be disconnected or can be pulled through.
Oxygen connection
Prior to ECMO initiation, the green oxygen tubing must be connected directly from the oxygen cylinder on the console to the oxygenator inlet. Ensure the white cap on the oxygenator inlet is removed before connecting the tubing. Air and oxygen “wall” connections must be established only after ECMO initiation, regardless of the mode or urgency of ECMO support. The gas outlet at the bottom of the oxygenator must not be occluded.
Running primed HLS circuit
Run the circuit prior to patient connection to ensure
- A functioning ECMO console
- Air-free circuit
- Correct pressure and flow monitoring
- Warming of primed circuit to body temperature (preferable)
The HLS system blood flow signal is obtained via an ultrasonic flow sensor clamped securely to the return (post oxygenator) line. The flow indicator arrow must point in the direction of circuit blood flow (away from the oxygenator).
- Remove yellow cap from oxygenator
- Turn console on, remove clamp
- Turn RPM up to 3000 RPM for 2 min
- Run at 4000 RPM for 1 min (look for light agitation in priming bag)
- Re-apply yellow cap
- Clamp circuit with ECMO clamp post oxygenator (also clamp both blue and red tubing at junction to reservoir bag with pre-attached white clamps)
- Prepare to hand over circuit for connection – open lid of sterile tray if connection is imminent
Rotaflow II start-up
Information on priming the PLS circuit can be found here. The video below shows the start-up.
Youtube video blocked
This YouTube video has been blocked from loading as you have not given consent to the associated cookies.
Rotaflow II Emergency Drive
An emergency drive unit (“hand crank”) must be sited on the trolley in a position whereby the centrifugal pump can be transferred immediately in the case of either console or pump head failure.
XXX rotaflow image
The emergency hand crank is freely accessible in proximity to the console, the heater connection may need to be disconnected or can be pulled through.
Oxygen connection
Prior to ECMO initiation, the green oxygen tubing must be connected directly from the oxygen cylinder on the console to the oxygenator inlet. Ensure that the white cap on the oxygenator inlet is removed before connecting the tubing. Air and oxygen “wall” connections must be established after ECMO initiation , regardless of the mode or urgency of ECMO support. The gas outlet at the bottom of the oxygenator must not be occluded.
Running primed PLS circuit
Running the circuit prior to patient connection to ensure
- A functioning ECMO console
- Air-free circuit
- Correct pressure and flow monitoring
- Warming of primed circuit to body temperature (preferable – see Heating the circuit)
The PLS system blood flow signal is obtained via an ultrasonic flow sensor clamped securely to the return (post oxygenator) line. The flow indicator arrow must point in the direction of circuit blood flow (away from the oxygenator).
- Remove yellow cap from oxygenator
- Turn console on, remove clamp
- Turn RPM up to 3000 RPM for 2 min
- Run at 4000 RPM for 1 min (look for light agitation in priming bag)
- Re-apply yellow cap
- Clamp circuit with ECMO clamp post oxygenator (also clamp both blue and red tubing at junction to reservoir bag with pre-attached white clamps)
- Prepare to hand over circuit for connection – open lid of sterile tray if connection is imminent