Console initial settings

Blood flow

Generally, the RPM is set to achieve the desired blood flow support in VA ECMO (e.g. achieving a CI >2.2 L/min/m2) and to achieve sufficient oxygenation in VV ECMO.

The maximum blood flow that can be delivered through the access cannula may be briefly assessed by increasing the pump speed in a stepwise manner up to 4000 RPM, unless access insufficiency occurs.

Sweep Gas flow

As a principal guide, the initial oxygen flow (sweep gas flow) in L/min should match the blood flow in L/min for VA and at the very least 1 L/min for VV ECMO. The initial aim is to prevent worsening of the acid-base status of the patient while avoiding rapid, large changes in arterial CO2 gas tension.

Patients in extremis, those with hypermetabolic states, and those with significantly increased minute ventilation before VV ECMO initiation may require higher starting oxygen flow rates e.g. 5 L/min.