Receive the tubing from the sterile tray and place clamps one hand width from the distal connection sites
- HLS: disconnect or cut at both ends within the tray
- PLS: receive both lines and cut out u-shaped connector
Connect the flushed access and return cannulae to the access and return limbs of the primed circuit. Ensure the access cannula is connected to access tubing (entering pump first, blue mark on tubing) and the return cannula is connected to return tubing (coming from oxygenator, red mark on tubing).
FINAL CHECKS – ‘BOC’
Mandatory checks by cannulator before connection – assistant to remind them
- [B] Bubbles – Check for air – tubing, oxygenator, pumphead
- [O] Oxygen – Oxygen connected and turned on
- [C] Clamp – Safety clamp on the circuit
Perform underwater seal connection
- Check connection to exclude air bubbles
- Cannulators remove clamps and state CLAMPS OFF
- Console operator sets pump at 1500 RPM* and removes clamp, then states: CLAMP OFF
- Steadily increase ECMO support (e.g. up to 3000-4000 RPM) to achieve desired blood flow
Confirmation
- Visualise direction of blood flow from the access cannula to pump head
- Bright red blood returning from the oxygenator confirms gas delivery and oxygenator function
Note
*Setting the RPM to 1500 is required in VA ECMO to overcome the pressure in the arterial system for VA ECMO.
Heating the circuit
Warming the circuit to body temperature before connection is mandatory in VV ECMO cannulation and desirable in VA ECMO if time permits (the risk of cardiac arrhythmias caused by infusing a large amount of cold saline rapidly is greatest in VV ECMO). In ECPR, pre-heating of the circuit may cause distraction or delay of ECMO commencement and should be done after the ECMO support is established when practical in the location. Also, the set temperature should not be higher than 36°C until the core temperature is known to avoid overshooting the target temperature.
For how to connect the heater watch the following video.