Cannula size, choice and position

Cannula size and choice

See table below for overview and cannula size guide as a starting point, specified according to clinical scenario and bedside judgement.

VV ECMO application

Multi-stage drainage cannula

23 Fr

Small patient/ partial support OR full support combined with 23 Fr return

25 Fr

Standard VV support

27 Fr

Consider for very large patient or 100% lung shunt

Single stage venous return cannula

19 Fr

Small patient or partial support (e.g. asthma)

21 Fr

Standard VV support (arterial cannula for fem-jug configuration)

23 Fr

Large patient

VA ECMO application

Multi-stage drainage cannula

19 Fr

ECPR

21 Fr

ECPR or elective VA small patient or partial support

23 Fr

Elective VA – standard normal size or large patient for full support

Single stage arterial return cannula*

15 Fr

ECPR/ elective VA small patient or partial support

17 Fr

ECPR/ elective VA most patients

19 Fr

Elective VA – full support in larger patient or CTH request

* Ultrasound measurement of artery diameter must be compatible with suggestion

Access cannula position

The optimal access cannula position is identical for VA and VV ECMO.