VAD Emergencies

Cardiac arrest

HM 3 shut down

This is only to be undertaken in the setting of palliation

VAD Controller/Pump

Cessation of Therapy

Controller will alarm during this process – unavoidable.

Place patients controller onto Battery power

(Not monitor or AC power)

Remove First Power Cable

Remove Driveline

Remove Second Power Cable

Press and hold the Battery button on Controller

Controller will shut down after 5 seconds of holding the button

Ensure Heartmate Console is plugged into power in the Dirty Respiratory Room once it is removed from the cubicle.

VAD Controller/Pump

Cessation of Therapy

Controller will alarm during this process – unavoidable.

1. Place patients controller onto Battery power

(Not monitor or AC power)

2. Remove First Power Cable

3. Remove Drive Line

4. Remove Second Power Cable

5. Press and hold the Battery button on Controller

6. Controller will shut down after 5 seconds of holding button.

Ensure Heartmate Console is plugged into power in the Dirty Respiratory Room once it is removed from the cubicle.